400 Series
The 430D Batch Controller is suited to applications where precise measurement and control of variable batch quantities is required


430D Batch Controller

Overview

The 430D Batch Controller is suited to applications where precise measurement and control of variable batch quantities is required.

The instrument is extremely flexible and easy to operate. Batches can be started, paused, stopped and reset via four operational keys and batch quantities and calibration parameters are entered via a 12 key numeric keypad.

Batch TotalFlow RateAccumulated Total and Preset Value can all be displayed in engineering units on the green LED display.

The 430D will accept most frequency and pulse signals, including mV outputs from turbine flowmeters, and two-wire proximity switch outputs.

Batch Features

Automatic Overrun Compensation Based on previous batches, the 430D will automatically compensate for any overrun which may be caused by slow closing valves.

Features at a glance

  • Accepts pulse and frequency flow signals
  • Scaled pulse output
  • Overrun compensation
  • One or two stage valve control
  • Optional RS232/422/485 interfaces for computers and printers
  • Remote start/stop
  • No-signal alarm
  • Ticket printing with time and date
  • Displays batch total, preset quantity, flow rate and accumulated total

430D Batch Controller

Computer/Printer Interface

An RS232 and RS422/485 option is available for interfacing to computers and printers. The interface allows the Batch Controller to be fully controlled from a computer.

Software drivers are included for a number of printers and can print a ticket with time and date, a sequential delivery/batch number and the batch and accumulated totals. An optional header for the ticket can also be printed.

Rate Totaliser Mode

The 430D can be configured during calibration to operate as a Rate Totaliser as distinct from a batch controller.

In this mode the RUN, STOP and Batch Set keys perform no function. The Rate key is used to toggle the display between Rate and Resettable Total.

By pressing and holding the Accum. key, whenever the instrument is not in data entry mode, the Accumulated Total will be displayed.

Versatile performance options

The instrument will operate from 14 to 28V dc or from 110 to 220V ac mains.

The facia is fully watertight to Nema 4X (IP65) and is resistant to most chemicals.

A total conversion feature enables the rate to be displayed in one engineering unit (eg. gallons/minute) and the totals to be displayed in another engineering unit (eg. barrels).

The standard instrument is supplied as panel mount. Field enclosures are optionally available.

Operation

Keypad operations are straightforward and four LED’s indicate the operational status at all times.

The unique keypad simplifies operation, making the instrument suited to use by operators with minimal training.

Four operational keys and a full numeric keypad on the front facia, enable the batch quantity to be set, and batch operations to be started, paused, stopped or reset.

Setting Batch Quantity

To set a batch quantity, the Batch Set key is pressed, and then the quantity is entered via the numeric keys. Once set, the Batch Set key is again pressed to return to Run mode.

The batch quantity can only be set while the instrument is in a non- operational state such as when the batch is complete or paused.

Once set, the batch quantity will be retained in non-volatile memory and will not alter until changed by the user.

To check the batch quantity while a batch is in progess press the Batch Set key.

Starting a Batch

To start the batch the RUN key is pressed. The Run LED will light and the instrument will begin to totalise from zero or, if programmed for the count down mode, the display will decrement from the batch quantity.

Stopping a Batch

The batch can be stopped at any time by pressing the STOP key. To continue the batch press the RUN key, or to abort the batch press the STOP key again.

When the batch has been interrupted the Pause LED will light. If the batch is aborted the Stop LED will light.

Resetting a Batch

The instrument can be programmed to reset in either of two ways.

By manually pressing the Reset key, at the end of a batch, the Batch Total will be reset to zero (if programmed to count up) or to the preset quantity (if programmed to count down).

If Auto Reset is programmed, however, the Batch Total will automatically reset when the RUN key is pressed.

One or two stage valve control

Two output relays provide control for single or dual stage valve operation. The first relay will energise at the start of the batch, and de-energise when the batch is complete.

The second relay can be programmed to energise at a set time after the start, and to de- energise at a set quantity prior to completion of the batch. This feature enables a slow startup and slow shutdown of the flow.

The Batch Controller can also be programmed to count up from zero or down from the batch quantity.

430D two stage valve control

Specifications

General

Display

6 digit 0.4” (10.2mm) high green LED

Display Update Rate

0.25s

Data Retention

All set-up parameters and totals are stored in a non-volatile memory with 10 years retention

Transducer Supply

8-24V dc field adjustable, 65mA max

Power Requirements

dc Supply:

14 to 28V dc, 300mA typical (no options)

ac Supply:

ac mains set internally to 95-135V ac or 190-260V ac

Operating Temperature

0 to 55°C

 

Relay Outputs

Maximum Power

1250VA

Maximum Voltage

250V ac, 30V dc

Maximum Current

5 Amps

 

Pulse Output

Pulse Width

10ms (negative going pulse).

Max Duty Cycle

49 pulses per second

Scaling

The pulse output is scaled and outputs one pulse each time the accumulated total increments

 

Approvals

Electrical

ETL (US) approved to UL508 and CSA

Interference

CE Compliance

Frequency Input

Frequency Range

Minimum:

0.25Hz on Rate, 0Hz on Total

Maximum:

10KHz

Scaling Range

0.100 to 50,000

Switching Threshold

2.5V (except for Namur proximity and coil input types)

Maximum Input Voltage

50V peak

Input Impedance

Logic Signal, CMOS, Pulse:

100KW

Open Collector or Reed Switch:

10KW

Namur Proximity (set dc out to 8v):

1KW

Switch or Reed Switch with debounce circuit (200Hz max):

10KW

Coil (20mC P-P min):

100KW

Coil (22mV P-P min):

2.4KW

 

Optional Enclosures

Field Enclosures

IP67 (Nema 4X)

 

Important: Specifications are subject to change without notice.

Dimensional Drawing

 

430D dimensional diagram

Terminal Descriptions


RS232/422/485 Options All Models
No.
20RS232 Signal Ground
21RS232 Data In
22RS232 Data Out
23RS422/485 (-) Data Out
24RS422/485 (+) Data Out
25RS422/485 (-) Data In
26RS422/485 (+) Data In
27RS232 CTS


Relay Outputs & Switches 430D
No.
28Not to be used
29Not to be used
30End of Batch/Pump Control Signal
31Relay 2 - Normally Open
32Relay 2 - Normally Closed
33Relay 2 - Common
34Relay 1 - Normally Open
35Relay 1 - Normally Closed
36Relay 1 - Common

Standard Terminal Connections 430D
No.No.
1Calibration Link8Flow Common (-)
2Signal Ground9Flow Pulse Input
3Remote RUN Switch10Pulse Out
4Remote Rate Switch11dc Power Out (8-24V)
5Remote Reset Switch12dc Ground
6Remote STOP Switch13dc Power Input
7Flow Alarm14Not to be used

Ordering Information

When specifying please indicate model(s) required using the following method

430D ordering

Product Manual Download