400 Series
The 405 Flow Computer can handle a wide range of flow applications, where high accuracy and flexibility are required


405

Overview

The 405 Flow Computer can handle a wide range of flow applications, where high accuracy and flexibility are required. It will interface to most flowmeters and versions are available for pulse and analog inputs.

Either the Flow Rate or a Resettable Totacan be continuously displayed in engineering units on the large six digit LCD display. A non-resettable Accumulated Total is also displayed whenever the DISPLAY key is pressed.

A scaled pulse output, suitable for driving remote totalisers is a standard feature and the instrument also provides an 8-24 Volts dc power supply for driving transducers.

RS232 and RS422/485 communications interface

Optional features include: an isolated 4-20mA output; a communications interface; and high/low flow rate alarms which are output on two Form C relays.

The communications option includes both RS232 and RS422/485 interfaces for communicating with a computer.

Software drivers are also included for a number of printers and can print a ticket with time and date, a sequential ticket number, and the resettable and accumulated totals.

Set-up data is stored in non- volatile memory

The 405 is initially set-up by following a calibration sequence that enables scaling factors, digital filtering and display formatting to be set and stored in a non-volatile memory which does not require battery backup.

The instrument will operate from 12 to 28V dc or from the 110/220V ac mains.

Each flow computer is supplied with a sheet of self-adhesive engineering unit labels. A recess, adjacent to the display, is provided in which the correct label can be placed.

Features at a glance

  • Large LCD display
  • Operates from 12-28V dc or ac mains
  • Scaled pulse output
  • Accepts 4-20mA and frequency flow inputs
  • Provides 8-24V dc transducer power
  • High accuracy
  • Self-adhesive engineering unit labels
  • Simplified programming
  • CE Compliant

Functions

  • Rate
  • Total (Resettable)
  • Accumulated Total

Options

  • 4-20mA isolated output
  • RS232 and RS422/485 communications interface
  • High/low alarm outputs

Digital filtering enhances Rate accuracy

Frequency fluctuations caused by pulsating flow through a flowmeter, often makes the Rate impossible to read with any precision.

The 405 has a digital filter which will average out these fluctuations and enable the Rate to be read to a four digit accuracy.

The degree of filtering is fully programmable which means that highly accurate and stable readings can be obtained without excessive lag.

Different mounting options available

The 405 is designed for panel mounting with the facia watertight to IP65 (Nema 4X) and resistant to most chemicals. A field enclosure and an explosion- proof enclosure are optionally available.

Models Available

Model 405A

An analog version with a fully isolated input. Inputs include 4-20mA, 0-20mA, 1-5V and 0-10V signals from flowmeters and pressure transducers. In addition to linear and square law input relationships, open channel applications can be handled via a programmable power on the input relationship.

Model 405B

A frequency input version that will accept pulse inputs and open collector inputs from opto-sensors or Hall effect devices.

Model 405D

Frequency input version with an input conditioning card capable of accepting mV signals from coils, two wire proximity sensors, reed switch inputs and most other pulse type signals.

Model 405LA

4-20mA temperature input version with temperature correction for general chemicals, LPG and petroleums to API tables (see separate data sheet on 405L and 414L).

Model 405LR

RTD temperature input version with temperature correction for general chemicals, LPG and petroleums to API tables (see separate data sheet on 405L and 414L).

Model 405Q

Frequency or pulse input with non- linearity correction and quadrature signal for measurement of bi-directional flows. The 405Q is highly suited to custody transfer applications where high accuracy and signal integrity is required.

Model 405S

For applications requiring the addition or subtraction of two flows this instrument will accept two frequency inputs and display Rate and Total for each channel.

Operation

The display of the 405 will normally show the Rate or Resettable Total (Net Total in the case of the 405LA, 405LR and 405S), as selected by the RATE orTOTAL keys on the keypad. An LED in the facia will indicate which function is currently displayed.

The DISPLAY key can be used to display the Accumulated Total. On the first press of the DISPLAY key, the display shows ACCTOT for one second followed by the actual total.

The Accumulated Total continuously totalises the flow and is not resettable via the front panel.

On reaching the maximum displayed total, all totals will roll over to zero and continue totalising. If, at any time, power is lost or the instrument is switched off, the totals will be stored in the non-volatile memory. When power is switched back on to the instrument, the stored totals will be recalled from memory and the totals will be incremented from the last values.

In the case of the Models 405LA and 405LR, if temperature compensation is selected, a second press of the DISPLAY key will show the product temperature or density if a density meter input is selected. A third press of the DISPLAY will show the actual Gross Total.

With the 405S, the first press of the DISPLAY key will show RATE 1, and the second press TOTAL 1, for input 1. The third press will show RATE 2, and the fourth press TOTAL 2, for input 2.

Specifications

General

Display

6 digit 0.7” (17.8mm) high LCD

Display Update Rate

0.25s.  (0.5s on 405A)

Data Retention

Set-up parameters and totals stored in non-volatile memory with 10 years retention

Decimal Points

The number of decimal points with which the rate and totals are displayed can be programmed

Total Conversion

It is possible to program a constant so that the rate can be displayed in one unit (eg. gal/m) and the total displayed with a different unit (eg. barrels)

Time Base

The rate can be displayed in units per second, minutes, hours or days

Transducer Supply

8-24V dc field adjustable, 50mA maximum

Power Requirements

dc Supply:

11.5-28.5V dc, 130mA typical current (no options)

ac Supply:

ac mains set internally to 95-135V ac or 190-260V ac

Operating Temperature

0 to 55ºC

Rear Connection

Terminal Block

Front Panel

Material:

Polycarbonate

Protection:

IP65 (Nema 4X)

Case:

Aluminium

Inputs

Analog Input (405A)

Type:

Isolated 4-20mA, 0-20mA, 1-5V and 0-10V

Input Impedance:

Current  250 ohms

Voltage   10k ohms

Accuracy 0.075%

Span:

0.1000 to 50,000

Zero:

0.00000 to 50,000

Cut-off Point:

A low flow rate cut-off can be programmed below which flow is not registered. The cut-off is programmed as a percentage of span

Relationship:

Linear, square root or programmable

Open Channel:

For open channel flowmeters the power of the input relationship is programmable between 0 and 9.99. With open channel selected, the polarity of the signal can also be programmed such that, 20mA represents maximum or minimum flow

Frequency (Pulse) Input

(405B, 405D, 405Q and 405S )

Minimum:

0Hz on Totals, 0.25Hz on Rate

Maximum:

10kHz for single input, 2.5 for quadrature

Input Circuits:

Will accept most sine logic and proximity switch inputs

Scaling Range:

0.1000 to 50,000

4-20mA Inputs

(Flow and Temperature in the Model 405LA – see separate data sheet.)

RTD Input

(Model 405LR – see separate data sheet.)

Outputs

Pulse Output

Not available for Model 405S

Type:

Open collector output with a pulse produced on each increment of the accumulated total

Maximum Rate:

49 pulses per second

Pulse Width:

10ms

Maximum Current:

Current sinking transistor output 100mA, 30V dc maximum

4-20mA Output

Not available for Model 405S

Type:

Fully isolated output corresponding to the displayed flow rate, suitable for driving a recorder or controller. Outputs available are 4-20mA, 0-20mA, 0-10V or 2-10V, with the minimum and maximum levels programmable

Resolution:

10 bits

Accuracy:

<0.05%

Maximum Load:

500 ohms from internal power, 950 ohms if externally powered

Communications Output

Type:

RS232, RS422 or RS485 interface is available for driving printers and communicating with computers

Baud Rate:

300 to 9600

Parity:

None, odd and even

Data Bits:

7 or 8

Protocols:

A number of protocols are included to interface to printers and computers

Time/Date:

A real time clock provides time/date printing on tickets

ID Code:

For multi-point communications, a unique address can be programmed

High/Low Flow

Type:

Two Form C relays provide a high and low flow rate alarm. Alarm points are programmable during set-up

Maximum Current:

5A

Maximum Voltage:

250V ac, 30V dc

Maximum Power:

1250VA

Approvals

Electrical

ETL (US) approved to UL 508 and CSA.

Interference

CE Compliance

Optional Enclosures

Field Enclosures

IP67 (Nema 4X)

Explosion Proof Enclosures

CENELEC, FM, CSA and SAA approved enclosures available for hazardous areas

Important: Specifications are subject to change without notice.

Contrec 405 Flow Measurement Computer diagram

Terminal Descriptions

Terminals Common to all models
No.
1Calibration
2Switch Common
11DC Power Out
12DC Ground
13DC Power In


RS232/422 Options All Models
No.
20RS232 Signal Ground
21RS232 Data In
22RS232 Data Out
23RS422/485 (-) Data Out
24RS422/485 (+) Data Out
25RS422/485 (-) Data In
26RS422/485 (+) Data In
27RS232 CTS


Isolated 4-20mA Output Option
No.
20Not Used
210 Volts
220-10 Volts Out
23-12 Volts
24I (-)
25I (+)
26+15 Volts
27Not Used


High & Low Relay Alarm Option
No.
31High - Normally Open
32High - Normally Closed
33High - Common
34Low - Normally Open
35Low - Normally Closed
36Low - Common

Terminals 28,29 & 30 on the option card are not used.

Terminals specific to each model

 

No.405A405B405D405LA405LR405Q & 405SNo.
3Rate SwitchNot UsedRate SwitchFlow Input Ch. 2Flow Input Ch. 2Flow Input Ch. 23
4Total SwitchNot UsedTotal SwitchNot UsedPT100 (+)Not Used4
5Reset SwitchNot UsedReset SwitchTemp. (4-20mA)PT100 Signal (+)Not Used5
6Program SwitchNot UsedProgram SwitchNot UsedPT100 Signal (-)Not Used6
7Not UsedNot UsedNot UsedFlow AlarmFlow AlarmFlow Alarm7
8Flow CommonFlow CommonFlow CommonFlow CommonFlow CommonFlow Common8
9Flow Signal (mA)Flow SignalFlow SignalFlow Input Ch. 1Flow Input Ch. 1Flow Input Ch. 19
10Pulse OutNot UsedPulse OutPulse OutPulse OutPulse Out10
14Flow Signals (Volts)Not UsedNot UsedFlow Input (4-20mA)PT100 (-)Not Used14

Product Manual Download